My question is on validity of the MPI instument say for the purpose of project is valid only if that posess a Calibration certificate by a TP agency.or otherwise is acceptable.
On Mon, Apr 19, 2010 at 12:54 PM, Bathula Raghuram (Mumbai - PIPING) <R.Bathula@ticb.com> wrote:
--In house is acceptable with a written calibration procedure. Permanent magnet type shall be calibrated ( check magnetic power) prior to use on daily base, and electromagnetic type on yearly basis.
From: materials-welding@googlegroups.com [mailto:materials-welding@googlegroups.com] On Behalf Of Nishad AK
Sent: Monday, April 19, 2010 2:06 PM
To: materials-welding@googlegroups.com
Subject: [MW:4868] Calibration of MPI yoke and validity.
Good day,I have a query regarding the calibartion of MPI yoke both Electomagnetic & permanent magnetic type.Does that should hold a Valid calibration certificate by the Third party calibration body or is that ok it has the manufactures certificate showing the weight lifting capacity and other test results.
--
Thanks and Regards
Nishad AK
To post to this group, send email to materials-welding@googlegroups.com
To unsubscribe from this group, send email to materials-welding-unsubscribe@googlegroups.com
For more options, visit this group's bolg at http://materials-welding.blogspot.com/
The views expressed/exchnaged in this group are members personel views and meant for educational purposes only, Users must take their own decisions w.r.t. applicable code/standard/contract documents.
--
Thanks and Regards
Nishad AK
To post to this group, send email to materials-welding@googlegroups.com
To unsubscribe from this group, send email to materials-welding-unsubscribe@googlegroups.com
For more options, visit this group's bolg at http://materials-welding.blogspot.com/
The views expressed/exchnaged in this group are members personel views and meant for educational purposes only, Users must take their own decisions w.r.t. applicable code/standard/contract documents.
No comments:
Post a Comment